Varian Model 979 Automatic Portable Leak Detector
Price:$12,500
ONE YEAR WARRANTY
FACTORY SERVICED
Turbo-pumped, Auto-calibrate, complete with cart, 11 CFM Varian DS302 Vacuum Pump, manual and one-year warranty. Factory serviced and ready for 120 volt operation, $12,500.
Just In! We are proud to include this Varian model 979 Portable Helium Leak Detector in our current inventory. This super clean unit has just undergone a factory service and is ready for your most demanding leak testing applications.
The 979 features a combination of Varian’s latest-generation electronics architecture with a unique multi-flow vacuum system, resulting in a very flexible but easy-to-use compact leak detector with extra-high sensitivity and pressure tolerance.
Usable in both evacuation mode and sniff-testing mode, the Model 979 is easy to use, thanks to a push-button front panel, which contains main leak-rate and pressure displays, dedicated operation buttons, and a touch-screen for setup and status readouts. The core machine is maintenance-free and features extensive protection from air inrushes, power failures, and possible operator errors.
Use of a large external primary pump results in fast pump-down times, while high test-port pressure tolerance reduces time required to reach test mode. Fully automated tuning, calibration (to NIST-traceable standard), and zeroing reduces likelihood of operator error and the Auto-Zero feature guarantees accurate measurements even in environments where the helium background is changing.
Technical Specifications
Sensitivity ranges in atmospheric-cc/sec helium
Std. sensitivity10-4 thru 10-9 High sensitivity10-5 thru 10-10
Minimum detectable leak at 1000 ppm ambient helium
Std. sensitivity5×10-10 atm-cc/sec helium High senstivity5×10-11 atm-cc/sec helium
Maximum test-port pressure 1 primary pump5 torr
Test-port connectorNW-25 KF flange (adapter to 1-1/8 in. compression fitting available)
Leak rate indication50-segment log or linear LED bar graph plus numeric readout on screen
Leak rate units displayedatm-cc/sec, mbar-l/sec, torr-l/sec, Pa-m³/sec
Test-cycle controlAutomatic
Calibration routineAutomatic or manual
Background suppression (zeroing)Automatic and Auto-Zero < Zero
Changing rangesAutomatic
Protection from air-inrush and power failureAutomatic
Audio signalBoth threshold and volume can be controlled
Remote analog leak-rate outputPhone jack, 0 to 5 VDC
Recommended ambient operating temperature
With rotary-vane pump(s)12 to 40 °C With dry scroll pump(s)5 to 40 °C
Power requirements115 V, 50 Hz | 115 V, 60 Hz | 230 V, 50/60 Hz (20 amps)
Weight (base unit only)22.7 kgs (50 lbs)